GC-FID

Gas Chromatography – Flame Ionization Detector or GC-FID is a very common analytical technique that is widely used in the petrochemical, pharmaceutical and natural gas markets. An FID typically uses a Hydrogen/Air flame into which the sample is passed to oxidise organic molecules and produces electrically charged particles (ions). The ions are collected and produce an electrical signal which is then measured. As common with other GC techniques, a carrier gas is required with low Water and Oxygen impurities since Water and Oxygen can interact with the stationary phase and cause significant problems such as high baseline noise and column bleed in the output gas chromatogram which both reduces the analyser sensitivity and decreases column lifetime. The FID is also extremely sensitive to Hydrocarbon impurities in the Hydrogen and Air supply for the flame. Hydrocarbon impurities can cause increased baseline noise and reduce the detector sensitivity.
 
 
 

 